Determination Method Validation of Copper in Electroplating Waste Water by American Public Health Association (APHA) 3111-B

Tugas Akhir, 010 / 2014 / ANK
Undergraduate Theses from JBPTPPOLBAN / 2018-03-29 14:50:50
Oleh : Lita Ayu Listiani - 111431016 & Octaviani Ratnasari - 111431023 (litaa.listiani@gmail.com; octaviani.ratnasari@yahoo.com)
Dibuat : 2018-03-29, dengan 4 file

Keyword : Validasi, Cu, Spektrofotometri Serapan Atom
Subjek : Validation, Cu, Atomic Absorption Spectrophotometer

Kontaminasi Cu dalam air limbah pada tingkat tertentu dapat menyebabkan pencemaran lingkungan dan gangguan kesehatan bagi makhluk hidup. Tujuan dari pekerjaan ini adalah untuk memvalidasi metode analisis Cu dalam limbah cair electroplating dengan menggunakan AAS. Pengerjaan validasi metode dilakukan berdasarkan APHA (American Public Health Association) Standard Method for Examination of Water and Waste Water22nd Edition nomer 3111-B. Parameter dalam validasi metode ini adalah linieritas, akurasi, presisi, batas deteksi dan batas kuantitasi. Berdasarkan pengukuran larutan standar didapatkan nilai linieritas sebesar 0,9997 dengan nilai regresi y = 0,126 x + 0,001. Nilai akurasi masing-masing 92.36; 92.92 dan 92.07%. Hasil untuk presisi repeatibility sebesar 1.81%, presisi reproducibility oleh lita 1.39% dan Octaviani 1.42%. Nilai batas deteksi dan limit kuantitasi masing-masing sebesar 0,3195 dan 0,1003. Secara keseluruhan hasil validasi untuk analisis Cu dalam air limbah dengan menggunakan Spektrofotometer Serapan Atom (AAS) dapat dikatakan baik, sehingga metode dapat digunakan untuk pelaksanaan analisis rutin di lingkungan laboratorium kimia "PT.SUCOFINDO Cabang Bandung". Kata Kunci: Validasi, Cu, Spektrofotometri Serapan Atom.

Deskripsi Alternatif :

The contamination of Cu in waste water at a certain level can cause environmental pollution and health problems for organism. The aim of this work is to validate a method of Cu analysis in waste water using AAS. The execution of method validation performed based APHA (American Public Health Association) Standard Method for Examination of Water and Waste Water22nd Edition number 3111-B. Parameters in the validation of this method are linearity, accuracy, precision, limits of detection and limits of quantitation. Based on measurement of standard the linearity was 0.9997 with a regression value of y = 0.126 x + 0.001. The value of accuration 92.36; 92.92 and 92.07% respectively. The result for repeatability precision is 1.81%, reproducibility precision by lita is 1.39% and Octaviani is 1.42%. The value of limit detection and limit of quantitation are 0.3195 and 0.1003 respectively. Overall the result of validation for Cu analysis in waste water using Atomic Absorption Spectrophotometer (AAS) are good, so the method can be use for implementation of routine analysis in chemistry environment laboratory PT. SUCOFINDO Bandung Subdivision". Keywords: Validation, Cu, Atomic Absorption Spectrophotometer.

Beri Komentar ?#(0) | Bookmark

PropertiNilai Properti
ID PublisherJBPTPPOLBAN
Organisasi
Nama KontakHelmi Purwanti
AlamatJl. Trsn. Gegerkalong Hilir Ds. Ciwaruga
KotaBandung
DaerahJawa Barat
NegaraIndonesia
Telepon022 201 3789 ext. 168, 169, 239
Fax022 201 3889
E-mail Administratorhelmi.purwanti@polban.ac.id
E-mail CKOhelmi.purwanti@polban.ac.id

Print ...

Kontributor...

  • Pembimbing: Dra. Nancy Siti Djenar, M.S.

    Pembimbing Lapangan: Drs. Faridz Mauladi, Editor: Erlin Arvelina

Download...